Aspiration-line pressure stability in Venturi and peristaltic pump configurations during phacoemulsification: an in vitro study

Tomoyuki Kunishige, Hisaharu Suzuki, Fumiki Okamoto

Short summary

Phacoemulsification using a Venturi pump configuration showed significantly more stable aspiration-line pressure (lower standard deviation) than a peristaltic pump configuration across tested vacuum settings (VAC200 and VAC400).

AI-generated from the title and abstract; the full text is not read.

Key points

  • Venturi pump configuration had lower mean aspiration-line pressure standard deviations (±0.53 mm Hg at VAC200, ±1.06 mm Hg at VAC400) than peristaltic pumps (±2.99 mm Hg at VAC200, ±2.54 mm Hg at VAC400).
  • Transient pressure fluctuations of 10 mm Hg or greater were observed only with the peristaltic pump at the VAC400 setting.
  • The study used an artificial anterior chamber model and two vacuum settings (VAC200, VAC400) to evaluate pressure waveforms.
  • Results represent within-run temporal fluctuation, not between-run variability.

AI-generated from the title and abstract; the full text is not read.

Abstract

This in vitro study descriptively evaluated aspiration-line pressure waveforms between Venturi and peristaltic pump configurations and their associated tubing systems using two Stellaris systems and a KITARO artificial anterior chamber model. VAC200 and VAC400 represented 200- and 400-mm Hg vacuum settings in the Venturi configuration and vacuum limits in the peristaltic configuration, with peristaltic flow settings of 20 and 35 mL/min, respectively. Pressure was recorded at 0.1-s intervals and analyzed during the stable 30- to 50-s period after activation. Only one continuous run was performed per condition. Reported standard deviations (SDs) represent within-run temporal fluctuation, not between-run variability. Mean pressure, SD, and transient fluctuations of 10 mm Hg or greater were evaluated. At VAC200, mean pressure was − 10.40 ± 0.53 mm Hg in the Venturi configuration and 11.02 ± 2.99 mm Hg in the peristaltic configuration. At VAC400, values were − 70.90 ± 1.06 and − 50.97 ± 2.54 mm Hg, respectively. Descriptive analysis suggested greater pressure fluctuation in the peristaltic configuration at both settings. Intermittent transient fluctuations of 10 mm Hg or greater occurred only in P-VAC400. These findings describe aspiration-line pressure waveforms under the tested in vitro conditions.

The authors' abstract, as published at the source. Scientific Reports, 2026 · DOI ↗