M-protein tespitinde gelişmiş hassasiyet ve özgüllük için bir LC-Q-TOF “HiRes Mass-fix” testinin doğrulanması

Validation of an LC-Q-TOF “HiRes Mass-fix” assay for improved sensitivity and specificity of M-protein detection

Ryan W. Pearce, Mindy Kohlhagen, Mark A. Martinez, Maria Alice V. Willrich ve diğerleri

Kısa özet

Sıvı kromatografisi kuadrupol zaman-uçuş kütle spektrometrisi (LC-Q-TOF MS) kullanan yeni bir "HiRes Mass-Fix" testi, önceki "Mass-Fix" yöntemine göre 58 numuneden 41'inde M-proteini tespit ederek kanda M-proteini tespitini iyileştirdi.

Yapay zekâ ile başlık ve abstract'tan üretildi; tam metin okunmaz.

Ana noktalar

  • "HiRes Mass-Fix" testi, hassasiyet, doğruluk, doğrusallık, özgüllük ve stabilite açısından analitik olarak doğrulanmıştır.
  • Test, ortalama 3.8 mg/L tespit limiti elde etmiştir.
  • "HiRes Mass-Fix", "Mass-Fix" yöntemiyle negatif olan 58 numuneden 41'inde M-proteini tespit etmiştir.
  • Kemik iliği NGF pozitifliğine kıyasla kan M-proteini pozitifliği %62'den %89'a yükselmiştir.
  • Kan-kemik iliği uyumu yeni test ile %71'den %58'e düşmüştür.

Yapay zekâ ile başlık ve abstract'tan üretildi; tam metin okunmaz.

Özet (abstract)

OBJECTIVES: To validate a blood-based liquid chromatography quadrupole time-of-flight mass spectrometry (MS) method "HiRes Mass-Fix" as a fit-for-purpose, patient-specific method for detecting low level M-proteins in patients who have reached negativity by MALDI-TOF MS "Mass-Fix". METHODS: Analytical validation included precision, accuracy, analytical measurement interval, limit of detection, specificity, ion suppression, carryover, and extracted-sample stability. Clinical performance was assessed by comparing 174 patients with bone marrow (BM) measurable residual disease (MRD) by next-generation flow cytometry (NGF) to both Mass-Fix and HiRes Mass-Fix. RESULTS: HiRes Mass-Fix demonstrated acceptable precision, accuracy, linearity, specificity, and stability for its intended use, with an average limit of detection of 3.8 mg/L. HiRes Mass-Fix detected M-proteins in 41 of 58 Mass-Fix negative samples. This raised blood MRD positivity from 62 to 89 % against a BM NGF positivity rate of 51 %. Blood-BM agreement fell from 71 % (kappa 0.43) by Mass-Fix to 58 % (kappa 0.15) by HiRes Mass-Fix, driven by blood-positive/BM-negative discordance. IgG was overrepresented among discordant cases (68 vs. 45 %, p=0.003). CONCLUSIONS: HiRes Mass-Fix was successfully validated and demonstrated to increase M-protein detection in samples negative by Mass-Fix. IgG recycling may explain the blood-positive/BM-negative discordance, and serial measurements will be needed to distinguish residual M-protein from ongoing production. Blood-based M-protein measurement is a less invasive, lower-cost alternative to bone marrow biopsy for ongoing patient monitoring.

Yazarların özeti; kaynağından alınmıştır. Clinical Chemistry and Laboratory Medicine (CCLM), 2026 · DOI ↗