Clinical Oral Implants Research· 2026Q1
Clinical Accuracy of Complete Arch Implant Scans Recorded by Using a Customized Metallic Ring‐Based Calibrated Framework: An In Vivo Study
- 0citations
- Q1SCImago
- 2026year
Short summary
A customized metallic ring-based calibrated framework (CF) achieved significantly higher trueness in complete arch implant scans (0.034 mm linear deviation) compared to a conventional scan body protocol (SB) (0.124 mm linear deviation), with a Q1 journal finding.
AI-generated from the title and abstract; the full text is not read.
Key points
- A calibrated framework (CF) achieved 0.034 mm mean linear deviation, significantly better than the conventional scan body (SB) protocol's 0.124 mm (p < 0.01).
- Angular deviations were also lower with the CF (0.31°) compared to SB (0.55°) (p < 0.01).
- Trueness was defined as deviation from a coordinate measuring machine reference, and precision as scan variability.
- The CF protocol showed greater trueness and lower variability in complete arch implant scans.
AI-generated from the title and abstract; the full text is not read.
Abstract
AIM: To evaluate and compare the positional accuracy of complete arch intraoral digital implant scans obtained using a conventional scan body protocol and a customized metallic ring-based calibrated framework. MATERIALS AND METHODS: A completely edentulous arch restored with six implants was scanned using two complete arch intraoral digital impression protocols: a conventional scan body protocol (SB) and a metallic ring-based calibrated framework (CF). Twenty scans were obtained per group using the same intraoral scanner. A coordinate measuring machine (CMM) dataset served as the reference standard. All STL files were imported into dental CAD software to design a standardized virtual implant-supported bar, which was subsequently analyzed using reverse engineering software. Linear inter-implant distances and angular deviations were calculated. Trueness was defined as the absolute deviation from the CMM reference, and precision as the variability among repeated scans. Between-group comparisons were performed using independent t-tests with Welch correction and Holm adjustment (α = 0.05). RESULTS: The mean linear deviation was 0.034 (0.004) mm for CF and 0.124 (0.019) mm for SB (p < 0.01). The mean angular deviation was 0.31 (0.05) degrees for CF and 0.55 (0.10) degrees for SB (p < 0.01). CONCLUSIONS: The calibrated splinting protocol showed greater trueness and lower variability compared with the conventional scan body approach.
The authors' abstract, as published at the source. Clinical Oral Implants Research, 2026 · DOI ↗
Continue with a free account
Ask the paper: 3 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.
Continue free on the webSign in with Google or Apple; no card needed. You come back to this paper.
On your phone:
Field: Oral Surgery
Oral SurgeryDentistry