Translational Vision Science & Technology· 2026Q1
Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography
- 0citations
- Q1SCImago
- 2026year
Short summary
Hyperparallel OCT (HP-OCT) and swept-source OCT (SS-OCT) show high agreement (ICCs ≥0.88) for anterior segment biometry, but are not interchangeable due to systematic and proportional biases, with HP-OCT generally measuring lower values.
AI-generated from the title and abstract; the full text is not read.
Key points
- HP-OCT and SS-OCT show high inter-device agreement (ICCs ≥0.88) for anterior segment biometry, except for iris thickness (ICC = 0.66).
- HP-OCT consistently measured lower values than SS-OCT for angle opening distance, trabecular-iris space area, iris curvature, iris thickness, and anterior chamber depth.
- Differences between devices increased with larger measurements, indicating proportional bias, particularly for angle opening distance and trabecular-iris space area.
- Despite high agreement, systematic and proportional biases limit the interchangeability of HP-OCT and SS-OCT measurements.
AI-generated from the title and abstract; the full text is not read.
Abstract
Purpose: To evaluate the inter-device agreement between hyperparallel optical coherence tomography (HP-OCT) and swept-source OCT (SS-OCT) in quantitative anterior segment parameters. Methods: Phakic subjects aged ≥40 years were consecutively recruited at a tertiary eye hospital. Horizontal B-scans (0°–180°) were analyzed. On SS-OCT, parameters were measured automatically; on HP-OCT, they were extracted manually using 3D Slicer. Agreement was evaluated using intraclass correlation coefficients (ICCs), Bland–Altman plots, and orthogonal regression. Results: A total of 118 sectors from 59 subjects were analyzed. ICCs were ≥0.88 for all parameters except iris thickness (IT) at 750 µm (IT750, ICC = 0.66). HP-OCT consistently measured lower values than SS-OCT for angle opening distance (AOD) at 500 µm (AOD500: –0.024 [–0.036 to –0.013]) and 750 µm (AOD750: –0.032 [–0.050 to –0.015]), trabecular–iris space area (TISA) at 500 µm (TISA500: –0.008 [–0.013 to –0.003]) and 750 µm (TISA750: –0.012 [–0.020 to –0.004]), iris curvature (IC: –0.015 [–0.021 to –0.009]), IT750 (–0.031 [–0.041 to –0.022]), and anterior chamber depth (–0.118 [–0.138 to –0.098]). For the following parameters, differences between devices increased with larger measurements, with HP-OCT yielding proportionally lower values (proportional bias): AOD500, AOD750, TISA500, TISA750, IC, and lens vault. Conclusions: For anterior segment biometry, HP-OCT and SS-OCT showed high ICCs but limited interchangeability because of systematic and proportional bias. Translational Relevance: Understanding how HP-OCT measurements compare with those from SS-OCT is essential for clinical integration, because inter-device agreement and interchangeability guide how HP-OCT can be applied and interpreted in practice.
The authors' abstract, as published at the source. Translational Vision Science & Technology, 2026 · DOI ↗