Trueness and Precision of Three Novel Scan Body Designs with Horizontal Extensions for Complete-Arch Implant Impressions: A Comparative In Vitro Study with Intraoral Scanning and Photogrammetry Systems

Üzeyir Arslan, Gülümser Evli̇oğlu, Erkan Sancaklı, Emrah Baca

Short summary

Three novel scan body designs with horizontal extensions, additively manufactured via vat photopolymerization, demonstrated comparable trueness and precision to original manufacturer scan bodies in complete-arch implant impressions, with all tested groups staying within the 100 µm linear deviation threshold.

AI-generated from the title and abstract; the full text is not read.

Key points

  • Three novel scan body designs with horizontal extensions were evaluated for trueness and precision in complete-arch implant impressions.
  • All tested groups, including novel designs, stayed within the 100 µm mean linear deviation threshold.
  • Mean linear deviations ranged from 31 ± 4 µm to 58 ± 18 µm, and mean angular deviations from 0.068 ± 0.012° to 0.538 ± 0.250°.
  • One photogrammetry system exceeded the 0.40° mean angular deviation threshold.

AI-generated from the title and abstract; the full text is not read.

Abstract

Accurate transfer of implant positions is critical for complete-arch implant-supported rehabilitations, yet intraoral scanning accuracy degrades over long edentulous spans, where errors accumulate. This in vitro study evaluated the trueness and precision of three novel auxiliary scan body designs with two horizontal extensions, additively manufactured by vat photopolymerisation, primarily against original manufacturer scan bodies and secondarily against a partial dentition scanning protocol and three photogrammetry systems (two extraoral, one intraoral). Four implants (Ø 3.8 × 10 mm) were placed in a prefabricated polyurethane edentulous maxillary model; anterior implants were parallel and perpendicular to the occlusal plane, while posterior implants were angled at 30°. Nine groups were assessed: one reference dataset and eight study groups, the latter each comprising 12 repeated scans. Deviations were calculated as differences in inter-implant centre distances (linear, µm) and inter-axis angles (angular, degrees) relative to a laboratory optical scanner. Group differences were significant for all endpoints (p < 0.001). Mean linear deviations ranged from 31 ± 4 µm to 58 ± 18 µm and mean angular deviations from 0.068 ± 0.012° to 0.538 ± 0.250°. No group exceeded the 100 µm mean linear threshold; one photogrammetry system exceeded the 0.40° mean angular threshold. Linear and angular results followed different rankings and should be evaluated separately.

The authors' abstract, as published at the source. Journal of Functional Biomaterials, 2026 · DOI ↗