Journal of Medical Engineering & Technology· 2014Q3
Comparative abilities of Microsoft Kinect and Vicon 3D motion capture for gait analysis
- 549citations
- Q3SCImago
- 2014year
Short summary
The Microsoft Kinect, when used with Brekel Kinect software, shows high correlation and small error for stride timing (r=0.94) but is not clinically accurate for hip and knee joint kinematics during gait analysis compared to Vicon Nexus.
AI-generated from the title and abstract; the full text is not read.
Key points
- Kinect and Vicon showed high correlation (r=0.94) and small error for stride timing measurements.
- Kinect generally under-estimated hip/knee flexion and over-estimated extension.
- Correlation between Kinect and Vicon for hip angular displacement was very low.
- Kinect knee measurements were somewhat better than hip but not clinically consistent.
- Kinect measurement variability was lowest at the slowest walking velocity.
AI-generated from the title and abstract; the full text is not read.
Abstract
Biomechanical analysis is a powerful tool in the evaluation of movement dysfunction in orthopaedic and neurologic populations. Three-dimensional (3D) motion capture systems are widely used, accurate systems, but are costly and not available in many clinical settings. The Microsoft Kinect™ has the potential to be used as an alternative low-cost motion analysis tool. The purpose of this study was to assess concurrent validity of the Kinect™ with Brekel Kinect software in comparison to Vicon Nexus during sagittal plane gait kinematics. Twenty healthy adults (nine male, 11 female) were tracked while walking and jogging at three velocities on a treadmill. Concurrent hip and knee peak flexion and extension and stride timing measurements were compared between Vicon and Kinect™. Although Kinect measurements were representative of normal gait, the Kinect™ generally under-estimated joint flexion and over-estimated extension. Kinect™ and Vicon hip angular displacement correlation was very low and error was large. Kinect™ knee measurements were somewhat better than hip, but were not consistent enough for clinical assessment. Correlation between Kinect™ and Vicon stride timing was high and error was fairly small. Variability in Kinect™ measurements was smallest at the slowest velocity. The Kinect™ has basic motion capture capabilities and with some minor adjustments will be an acceptable tool to measure stride timing, but sophisticated advances in software and hardware are necessary to improve Kinect™ sensitivity before it can be implemented for clinical use.
The authors' abstract, as published at the source. Journal of Medical Engineering & Technology, 2014 · DOI ↗
Continue with a free account
Ask the paper: 3 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.
Continue free on the webSign in with Google or Apple; no card needed. You come back to this paper.
On your phone:
Field: Physical Therapy, Sports Therapy and Rehabilitation
Physical Therapy, Sports Therapy and RehabilitationHealth Professions