Annals of Laboratory Medicine· 2026Q1
Application of Patient-Based p Control Chart for Quality Control in a Clinical Laboratory
- 0citations
- Q1SCImago
- 2026year
Short summary
A patient-based p control chart effectively complements traditional quality control in clinical laboratories, demonstrating low false-positive rates (0.25%-0.50%) and successfully identifying subtle shifts in assays like HbA1c and total protein that internal QC missed.
AI-generated from the title and abstract; the full text is not read.
Key points
- Patient-based p control charts offer a complementary QC method for clinical laboratories.
- Empirical false-positive rates were low, ranging from 0.25% to 0.50% under in-control conditions.
- The p chart detected subtle, sustained shifts in HbA1c and total protein results missed by internal QC.
- Chart performance is sensitive to positive rate, threshold location, and subgroup size.
- Greater stabilization of false-positive rates was achieved with daily subgroup sizes of 200 or more.
AI-generated from the title and abstract; the full text is not read.
Abstract
Background: Patient-based QC is a critical complementary tool for laboratory quality management. We evaluated the application and statistical performance of a patient-based p (a proportion-based statistical control chart) across quantitative and qualitative laboratory assays. Methods: We collected 413,586 valid test records involving 10 laboratory assays and divided them into training and verification datasets. P control charts were constructed using daily positive proportions and sample-size-specific limits. Monte Carlo simulations evaluated in-control performance, shift-detection performance, and sensitivity to group size using theoretical binomial shifts, empirical bootstrap simulations with analytical bias, and fixed subgroup sizes. Results: Empirical false-positive rates under in-control conditions ranged from 0.25% to 0.50%. Detection power increased with the magnitude of event-proportion or analytical shifts, whereas directional asymmetry occurred for thresholds located in distribution tails. Sample-size simulations showed greater stabilization of false-positive rates at daily subgroup sizes of ≥ 200. With clinical monitoring, the p control chart identified sustained increases in HbA1c positivity exceeding the diagnostic cut-off and detected persistent changes in total protein patient-result proportions that were not accompanied by conventional internal QC violations. Clinical decision limits-based charts for Na and K generated fewer signals than did reference intervals-based charts. In HBsAg monitoring, the p control chart remained within control limits following reagent lot replacement, whereas the corresponding Levey–Jennings charts showed an approximately 4-SD shift. Conclusions: The patient-based p control chart complemented laboratory QC analysis, with performance influenced by positive rate, threshold location, and subgroup size.
The authors' abstract, as published at the source. Annals of Laboratory Medicine, 2026 · DOI ↗
Continue with a free account
Ask the paper: 5 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.
Continue free on the webSign in with Google or Apple; no card needed. You come back to this paper.
On your phone:
Medical Laboratory TechnologyHealth Professions