European Journal of Nuclear Medicine and Molecular Imaging· 2014Q1
FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0
- 3,310citations
- Q1SCImago
- 2014year
Short summary
These updated guidelines provide standardized procedures for FDG PET/CT in adult oncological imaging, focusing on harmonizing quantitative measurements for improved accuracy and reproducibility across different sites and systems.
AI-generated from the title and abstract; the full text is not read.
Key points
- Provides standardized procedures for FDG PET/CT in adult oncological imaging.
- Focuses on quality control and assurance for accurate and precise quantitative measurements.
- Aims to achieve harmonization of standardized uptake value (SUV) across multicenter settings.
- Enhances repeatability and reproducibility of imaging results.
AI-generated from the title and abstract; the full text is not read.
Abstract
The purpose of these guidelines is to assist physicians in recommending, performing, interpreting and reporting the results of FDG PET/CT for oncological imaging of adult patients. PET is a quantitative imaging technique and therefore requires a common quality control (QC)/quality assurance (QA) procedure to maintain the accuracy and precision of quantitation. Repeatability and reproducibility are two essential requirements for any quantitative measurement and/or imaging biomarker. Repeatability relates to the uncertainty in obtaining the same result in the same patient when he or she is examined more than once on the same system. However, imaging biomarkers should also have adequate reproducibility, i.e. the ability to yield the same result in the same patient when that patient is examined on different systems and at different imaging sites. Adequate repeatability and reproducibility are essential for the clinical management of patients and the use of FDG PET/CT within multicentre trials. A common standardised imaging procedure will help promote the appropriate use of FDG PET/CT imaging and increase the value of publications and, therefore, their contribution to evidence-based medicine. Moreover, consistency in numerical values between platforms and institutes that acquire the data will potentially enhance the role of semiquantitative and quantitative image interpretation. Precision and accuracy are additionally important as FDG PET/CT is used to evaluate tumour response as well as for diagnosis, prognosis and staging. Therefore both the previous and these new guidelines specifically aim to achieve standardised uptake value harmonisation in multicentre settings.
The authors' abstract, as published at the source. European Journal of Nuclear Medicine and Molecular Imaging, 2014 · DOI ↗