Dentomaxillofacial Radiology· 2014Q1
Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units
- 478citations
- Q1SCImago
- 2014year
Short summary
Dental CBCT effective doses vary widely (adults: 5-1073 µSv; children: 7-769 µSv) depending on field of view and unit, with dose-height product (DHP) reducing calculation error by 45% compared to dose-area product (DAP).
AI-generated from the title and abstract; the full text is not read.
Key points
- Adult effective doses for dental CBCT range from 5 to 1073 µSv across different field of view (FOV) sizes.
- Child effective doses range from 7 to 769 µSv for large/medium FOVs and 7 to 521 µSv for small FOVs.
- Mean adult effective doses were 212 µSv (large FOV), 177 µSv (medium FOV), and 84 µSv (small FOV).
- Dose-height product (DHP) reduced average absolute error in dose calculation by 45% compared to dose-area product (DAP).
AI-generated from the title and abstract; the full text is not read.
Abstract
OBJECTIVES: This article analyses dose measurement and effective dose estimation of dental CBCT examinations. Challenges to accurate calculation of dose are discussed and the use of dose-height product (DHP) as an alternative to dose-area product (DAP) is explored. METHODS: The English literature on effective dose was reviewed. Data from these studies together with additional data for nine CBCT units were analysed. Descriptive statistics, ANOVA and paired analysis are used to characterize the data. RESULTS: PubMed and EMBASE searches yielded 519 and 743 publications, respectively, which were reduced to 20 following review. Reported adult effective doses for any protocol ranged from 46 to 1073 µSv for large fields of view (FOVs), 9-560 µSv for medium FOVs and 5-652 µSv for small FOVs. Child effective doses from any protocol ranged from 13 to 769 µSv for large or medium FOVs and 7-521 µSv for small FOVs. Effective doses from standard or default exposure protocols were available for 167 adult and 52 child exposures. Mean adult effective doses grouped by FOV size were 212 µSv (large), 177 µSv (medium) and 84 µSv (small). Mean child doses were 175 µSv (combined large and medium) and 103 µSv (small). Large differences were seen between different CBCT units. Additional low-dose and high-definition protocols available for many units extend the range of doses. DHP was found to reduce average absolute error for calculation of dose by 45% in comparison with DAP. CONCLUSIONS: Large exposure ranges make CBCT doses difficult to generalize. Use of DHP as a metric for estimating effective dose warrants further investigation.
The authors' abstract, as published at the source. Dentomaxillofacial Radiology, 2014 · DOI ↗
Continue with a free account
Ask the paper: 3 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.
Continue free on the webSign in with Google or Apple; no card needed. You come back to this paper.
On your phone:
Field: Oral Surgery
Oral SurgeryDentistry