Quantification of the flexural rigidity of endovascular surgical devices using three-point bending tests
- 9citations
- Q1SCImago
- 2026year
Short summary
A new three-point bending test methodology quantifies lengthwise flexural rigidity profiles of endovascular devices, revealing distinct property ranges for guidewires, intermediate guides, and long sheaths.
AI-generated from the title and abstract; the full text is not read.
Key points
- Developed a repeatable three-point bending test methodology for quantifying lengthwise flexural rigidity of endovascular devices.
- The methodology can characterize transition zones in device mechanical properties.
- Generated the first comprehensive quantitative library of flexural rigidities for guidewires, intermediate guides, and long sheaths.
- Observed distinct flexural rigidity ranges for the three device classes.
- Identified application-specific trends between flexural rigidity, device diameter, and length.
AI-generated from the title and abstract; the full text is not read.
Abstract
Endovascular surgical procedures require the navigation of catheters and wires through the vasculature to reach distal target sites. Quantitative frameworks for device selection hold the potential to improve the tracking of endovascular devices through vascular anatomy by personalizing the device flexural rigidity to an individual's anatomy. However, data are lacking to facilitate this technology, in part because typical endovascular devices have intricate spatial variations in mechanical properties that are challenging and tedious to quantify repeatably. We therefore developed a three-point bend test methodology using a custom rig and applied it to measure lengthwise flexural rigidity profiles of endovascular devices that are used to target the cerebral vasculature. The methodology demonstrated high repeatability and was able to characterize transition zones. We applied the methodology to generate the first comprehensive, quantitative library of device flexural rigidities, spanning guidewires, intermediate guides, and long sheaths. We observed that these three classes of device have properties that fall into distinct ranges. Additional plots examining relationships between flexural rigidity, device diameter, and length revealed application-specific trends in flexural properties. This methodology and the data allow for standardized characterization and comparisons to aid device selection, and have the potential to both enhance surgical planning and inform future innovation.
The authors' abstract, as published at the source. Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials, 2026 · DOI ↗
Continue with a free account
Ask the paper: 5 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.
Continue free on the webSign in with Google or Apple; no card needed. You come back to this paper.
On your phone:
Field: Biomedical Engineering
Biomedical EngineeringEngineering