Analytical Chemistry· 2026Q1
Apparent pH Sensitivity of Non-Functionalized Graphene FETs Originates from Faradaic Reactions at the Gold Gate
- 0citations
- Q1SCImago
- 2026year
Short summary
The apparent pH sensitivity of non-functionalized graphene field-effect transistors (GFETs) with gold gate electrodes is caused by Faradaic reactions at the gold gate, not the graphene itself.
AI-generated from the title and abstract; the full text is not read.
Key points
- Graphene FETs (GFETs) with gold gate electrodes show apparent pH sensitivity, while those with Ag|AgCl electrodes do not.
- The observed pH sensitivity (-62.7 mV pH−1) is linked to Faradaic reactions at the gold gate, not the graphene channel.
- Cyclic voltammetry and chronoamperometry distinguish charging current from Faradaic contributions.
- The gold gate itself is pH sensitive and its potential relaxes slowly.
AI-generated from the title and abstract; the full text is not read.
Abstract
Abstract Graphene field-effect transistors (GFETs) with solid-state gold gate electrodes have recently been used for accurate chemical analysis in complex matrices, yet the mechanistic origin of their pH sensitivity is not fully understood. Here we show that replacing only the gate electrode transforms an otherwise pH-insensitive graphene device into a pH-sensitive one. Using the same high-quality, low-defect CVD graphene as the channel, devices gated with a conventional Ag|AgCl reference electrode exhibit negligible pH sensitivity, while gold-gated devices show a response of −62.7 mV pH−1 at a gate voltage scan rate of 0.3 V s−1. Chronoamperometry distinguishes the transient charging current from the subsequent Faradaic contribution. Cyclic voltammetry and scan-rate-dependent measurements support a coupled Faradaic mechanism. The gate leakage current scales with the square root of the gate voltage scan rate (v), whereas the Dirac-point voltage varies approximately linearly with v1/2 and has a non-zero intercept. Independent open-circuit potential (OCP) measurements confirm that the Au gate itself is pH sensitive and relaxes slowly toward a stable potential. The GFET sensitivity decreases in magnitude at lower scan rates, and extrapolation as the scan rate approaches zero brings it closer to the OCP sensitivity. The remaining difference may reflect the different electrochemical conditions at the Au gate under bias and at open circuit. These findings indicate that the apparent pH response in Au-gated GFETs originates from gate-electrode electrochemistry and is read out through field-effect modulation of the graphene channel.
The authors' abstract, as published at the source. Analytical Chemistry, 2026 · DOI ↗
Continue with a free account
Ask the paper: 5 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.
Continue free on the webSign in with Google or Apple; no card needed. You come back to this paper.
On your phone:
Field: Analytical Chemistry
Analytical ChemistryChemistry