An Autonomous Robotic Phlebotomy Device: Extended Analytical Validation and Real-World Hemolysis Assessment

Jeroen A Roest, Luuk Giesen, Thijs C. van Holten, Martine M. L. Deckers et al.

Short summary

An autonomous robotic phlebotomy device (ARPD) demonstrated analytical equivalence to manual blood collection across different tube types and analytes, and significantly reduced hemolysis rates in real-world clinical settings, achieving 0.0% hemolysis compared to 1.6% for manual collection.

AI-generated from the title and abstract; the full text is not read.

Key points

  • The Aletta Autonomous Robotic Phlebotomy Device (ARPD) showed analytical equivalence to manual blood collection for key analytes and tube types.
  • Hemolysis index was statistically lower in ARPD-collected lithium heparin samples in one cohort.
  • In real-world use, the ARPD achieved a 0.0% hemolysis rate across 1368 venipunctures.
  • The ARPD's real-world hemolysis rate was significantly lower than manual phlebotomy (1.6%, p = 0.025).

AI-generated from the title and abstract; the full text is not read.

Abstract

Background/Objectives: Automated phlebotomy may improve preanalytical quality and access to venous blood testing. Prior work evaluating the Aletta Autonomous Robotic Phlebotomy Device (ARPD) demonstrated analytical equivalence. This study expanded the initial analytical validation to additional blood collection tubes and analytes and also assessed real-world hemolysis rates. Methods: The study was conducted in three hospital outpatient phlebotomy departments, including a site with a high prevalence of (hemato-)oncology patients. The within-subject analytical validation study included Greiner VACUETTE (Cohort 1) and BD Vacutainer (Cohort 2) serum, lithium heparin, and EDTA tubes. Equivalence of potassium, lactate dehydrogenase, hemolysis index, and complete blood count (CBC) with differential between paired ARPD-collected and manually collected samples was assessed using Passing–Bablok regression. Real-world ARPD hemolysis (>0.5 g/L free hemoglobin) rates were determined in a separate cohort (Cohort 3) and compared with site-specific manual hemolysis rates. Results: In the analytical validation study, 119 patients were included in the analysis (Cohort 1, n = 52; Cohort 2, n = 67). All tests met the predefined acceptability criteria. The hemolysis index was statistically significantly lower in ARPD-collected lithium heparin samples in Cohort 1. In real-world use (1368 ARPD venipunctures), the ARPD hemolysis rate was 0.0% (95% CI: 0.0–0.3%), significantly lower than manual phlebotomy (1.6%, p = 0.025). Conclusions: The study demonstrated analytical equivalence in a clinically complex population and found reduced ARPD hemolysis rates. The results further substantiate that the ARPD could improve patient care by reducing preanalytical errors.

The authors' abstract, as published at the source. Diagnostics, 2026 · DOI ↗