PofoliaDaily literature tracker

Journal of Applied Clinical Medical Physics· 2026Q1

Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance

Miia Nadhum, Anna Vuorenmaa, Ullamari Hakulinen, Antti Vehkaoja

Short summary

Semiautomated ultrasound quality assurance (QA) using UltraIQ software provides depth of visualization (DoV) measurements that are 5.0%-14.5% deeper than visual methods, while maintaining similar repeatability and reliability.

AI-generated from the title and abstract; the full text is not read.

Key points

  • Semiautomated analysis with UltraIQ provided deeper DoV measurements (5.0%-14.5%) compared to visual methods.
  • Visual DoV on PACS using uniformity images showed the closest agreement with UltraIQ results.
  • Both visual and semiautomated methods demonstrated good to excellent reliability (ICC 0.882-0.939).
  • UltraIQ offers a more objective approach to ultrasound QA.

AI-generated from the title and abstract; the full text is not read.

Abstract

BACKGROUND: Depth of visualization (DoV) measurements are recommended as part of routine ultrasound (US) quality assurance (QA) in international guidelines. Effective QA requires measurement and analysis procedures that are repeatable, reliable, and minimally operator‑dependent. PURPOSE: This study compares three visual DoV analysis methods with semiautomated analysis tool UltraIQ to identify the approach with the highest repeatability and improved objectivity. METHODS: A total of 215 retrospective DoV measurements acquired between 2018-2022 were analyzed, including 40 transducer models from 14 US devices by six manufacturers. Visual DoV measurements were performed on the US system display and on a medical-grade display using a Picture Archiving and Communication System (PACS) using the vertical distance group and the uniformity images obtained from QA Phantom. DoV was also evaluated using SNR-based analysis in the UltraIQ software. Repeatability was evaluated using three repeated DoV measurements for 17 transducers, and reliability was assessed using the intraclass correlation coefficient (ICC). RESULTS: UltraIQ provided DoV values that were 5.0%-14.5% deeper than those obtained with visual methods. Visual DoV on the US device display was 0.7-2.0 cm shallower than UltraIQ and 0.1-0.9 cm shallower than visual DoV on PACS. Visual measurements using uniformity images on PACS were 0.4-1.1 cm shallower than UltraIQ but demonstrated the closest agreement with UltraIQ results. ICC analysis showed good to excellent reliability. For visually measured DoV, ICC values ranged 0.893-0.939, while UltraIQ provided an ICC of 0.882. CONCLUSION: Semiautomated DoV analysis using UltraIQ is as repeatable and reliable as visual assessment while providing a more objective approach for US QA.

The authors' abstract, as published at the source. Journal of Applied Clinical Medical Physics, 2026 · DOI ↗