PofoliaShared via Pofolia

npj Digital Medicine· 2026Q1

Predicting calcification risk in prosthetic aortic valves: a hybrid physics-based and machine learning approach

Pascal Corso, Giorgia Tagliavini, Maria Giuseppina Chiara Nestola, Fergal B. Coulter

Short summary

A hybrid physics-based and machine learning model predicts calcification risk in prosthetic aortic valves, achieving a Spearman correlation of ρ S ≥0.9 against experimental data.

AI-generated from the title and abstract; the full text is not read.

Key points

  • A hybrid model combines fluid-structure interaction simulations with Finite-Time Lyapunov Exponents (FTLE) and wall shear stress (WSS) to predict calcification risk in prosthetic aortic valves.
  • The model partitions leaflets into four calcification risk classes using unsupervised k-means clustering, without fitting to experimental data.
  • The combined FTLE and WSS classification achieved a Spearman correlation ρ S ≥0.9 when validated against an experimental micro-CT calcification intensity map.
  • Different valve materials exhibit distinct strain and shear behaviors: porcine tissue localizes strain, while elastomers show flutter.

AI-generated from the title and abstract; the full text is not read.

Abstract

Abstract Calcific degeneration remains the foremost failure mode of bioprosthetic aortic valves and a mounting concern for polymeric substitutes. Six prosthetic valve configurations, pairing two leaflet scallop geometries (V-shaped D1 V and U-shaped D2 U ) with three constitutive materials (elastomer, bovine pericardium and porcine tissue), are simulated throughout systole within a patient-specific curved aorta reconstructed from clinical CT, using a high-fidelity fluid-structure interaction framework. An incremental formulation of Finite-Time Lyapunov Exponents (FTLEs), applied to the leaflet deformation gradients, is introduced as a descriptor of leaflet structural point coherence. With wall shear stress (WSS) derived features, the FTLE field feeds an unsupervised k -means clustering that partitions each leaflet into four calcification risk classes with no fitting to experimental data. Validated against an experimentally calibrated micro-CT calcification intensity map of explanted bovine pericardial leaflets, the combined FTLE and WSS classification attains a Spearman correlation ρ S ≥0.9 under near-equal weighting ( γ ≈ 0.5). Porcine tissue concentrates strain into localised high-gradient zones whereas the isotropic elastomer spreads strain evenly yet sustains pronounced shear fluctuations through leaflet flutter. The D2 U geometry with bovine pericardium proves least susceptible across both risk dimensions, furnishing a transferable, model-free tool for ranking next-generation valve designs.

The authors' abstract, as published at the source. npj Digital Medicine, 2026 · DOI ↗

TakeawaysPremium
Ask the paperFree account

Continue with a free account

Ask the paper: 3 free questions a day about this paper; save it, get its citation, new summaries every day for your field. Takeaways are Premium.

Continue free on the web

Sign in with Google or Apple; no card needed. You come back to this paper.

On your phone:

Field: Cardiology and Cardiovascular Medicine

Cardiology and Cardiovascular MedicineMedicine